Description
Dimesions
- Width: 280mm (5.9in)
- Height:152 mm (11in)
- Depth: 368 mm (14.5 in)
- Weight: 8 kg (17.6 lbs)
Environmental conditions for shipping and storage of the unit
- Temperature: -40°C to + 70°C
- Air Humidity, reative: 10% to 95%
Environmental conditions for operation of the unit
- Temperature: +10°C to + 40°C
- Air Humidity, reative: 130% to 75%, noncondensing
Power Conditions
- Rated power voltage: 240 V / 230 V / 115 V / 110 V / 100 V ± 10 %
- Rated power frequency: 50 / 60 Hz
- Power Current: 3.0 A at 230 – 240 V / 6.0 A at 100 – 115 V
- Power consumption in Standby mode: 25 Watts
- Power consumption at max. HF power: 450 Watts 690 VA
- Current consumption in Standby mode: 150 mA at 230 – 240 V / 300 mA at 100 – 115 V
- Potential equalization terminal: Yes
- Power Fuse: 2 slow burn, 4 A at 230 – 240 V / 8 A at 100 – 115 V
Energy
- Max. energy output ( at rated ohms) (Systems) Up to 250 Watt 200
- Max voltage, vp-p (Systems) Software controlled
- Max. energy output ( at rated ohms) (Systems) Up to 150 Watt 120
- Max voltage, vp-p (Systems) 380 soft, 5,600 spray
- Bipolar (Systems) Coagulate
- Max. energy output ( at rated ohms) (Systems) Up to 150 Watt 120
- Max voltage, vp-p (Systems) 380
Cutting/ AutoCut with Automatic Voltage control
- HF voltage waveform unmodulated sinusoidal alternating voltage
- Crest factor C, at RL=500 ohms C = 1.4 for all settings
- Rated frequency 330 kHz Maximum HF voltage at RL=∞ 600 Vp
- Dynamic internal impedance in the rated load range 0 ohms
- Cutting quality 4 coagulation effects selectable
- Consistancy of 4 coagulation effects automatically controlled
- HF rated power 200 watts at RL = 500 ohms H
- HF power limitation 1 to 200 watts in 1 watt steps
- Setting the HF power limitation via up/down keys
- Display of HF power limitation 7-segment display, 3 decimal places
- Precision of the HF power limitation +/- 1 digit or +/- 15%
- Automatic power control (PPS) yes Activation of the cutting mode via key or via pedal
Soft Coagulation
- HF voltage waveform unmodulated sinusoidal alternating voltage
- Crest factor C, at RL = 500 ohms C = 1.4 for all setttings
- Rated frequency of the HF voltage 330 kHz
- Peak value of the HF voltage max. 190 Vp
- Constancy of the HF voltage automatically controlled up to PHFmax.
- HF rated power 120 watts at 125 ohms
- HF power limitation (PHFmax.) from 1 watt to 120 watts in 1 watt steps
- Setting of the HF power limitation via up/down keys
- Display of the HF power limitation 7-segment display, 3 decimal places
- Precision of the HF power limitation +/- 1 digit or +/- 15%
- Activation of Soft coagulation via key or via pedal
Forced Coagulation
- HF voltage waveform pulse-modulated alternating current (ST generator)
- Crest factor C, at RL=500 ohms C = 5 at 120 Wmax, C = 11 at 5 Wmax
- Rated frequency of the HF voltage 1 MHz
- Peak value of the HF voltage, Version 1 max. 1300 Vp
- Peak value of the HF voltage, Version 2 max. 2300 Vp
- Peak value of the HF voltage, Version 3 max. 2300 Vp
- HF rated power 120 watts at 350 ohms
- HF power limitation from 1 watt to 120 watts in 1 watt steps
- Setting of the HF power limitation via up/down keys
- Stability of the HF power see diagram Forc. Coag. power via RL
- Display of the HF power limitation 7-segment display, 3 decimal places
- Precision of the HF power limitation +/- 1 digit or +/- 15%
- Activation of the Forced coagulation via key or via pedal
For more information, please feel free to contact us here!